Infusion Pump WIT-601A
- General Details
More Intelligent : Six Infusion Mode / BOLUS / History Record / Drug Library / Programming Infusion / PIGGYBACK
More Convenient: Open System / Digital Keyborad and Menu / Flexiable Fixation / Battery discharge time more than 7 hours
Much Safer: Double CPU / Lock Key / Anti-Reverse Function
More Comforable: Fluid Warmer Relieve unconfortableness
- Features & Specification
IV Set Options /Size Open System with IV set of 10d/ml、15d/ml、20d/ml、60d/ml(3.4mm~4.5mm); Also can be customized for Dedicated IV set;
Infusion Mode Volume Mode / Drop Mode / Time Mode /Programming Mode/Weight mode /DrugLiabrary Mode
Accuracy Error Rate ±5%
Flow Rate Range 1~1200ml/h(20 d/ml IV set)/ 1~150ml/h(60 d/ml IV set)
Drip Rate Range 1~400d/min (20d/ml IV set) / 1~150d/min (60d/ml IV set)
Preset Volume Range 1-9999ml
KVO 1ml/h
Purge 600ml/h
Bolus Default 600ml/h with volume 2ml;Other value can be setted
Bubble Detection Detected by ultrasonic with adjustive setting , 50 ~ 800μL.
Occlusion Pressure Upper position detected;Down position:High 70-90KPA;Mid 55-70KPA;Low 40-55KPA;
Alarm Bubbles, Door open, Finished, Pause Overtime,Leakage, Empty , Battery undervoltage or damage, Motor reverse, Incorrect IV Set-type setting, Pressure test failure, Motor failure,Abnormal drip sensor /drip rate, Heater failure, Down Occlusion,Up Occlusion,internal communication fail
Anti-reversal Infusion Anti-reverse function of motor to prevent upstream
CPU Dual
Lock Key Automatic Locking to avoid misoperation.
Programmable Infusion can be programmed in 8 segment
Library User can edit 100 kinds of drug
History Record 1500 event records , 200 modified recordsRecords can be exported in LAN
Piggyback User change parameters without switching off
Fluid Warmer(Selective) Relieve unconfortableness such as pain, vessel convulsion ,shiver when transfuse cold fluid
Battery Discharge time:Appr. 7 hours ; Recharge time: 12 hours
Rechargeable Ni-MH battery: DC9.6V, 2000mAh; Vehical-mounted adapter for ambulance(Selective)
Power Supply AC: 100V ~ 240V, 50Hz/60Hz; Power dissipation :25VA
Dimension 130mm×120mm×195mm
Weight 2.0Kg
Operation Condition Temperature: 5°C ~ 40°C;Relative humidity: 20% ~ 90%
Atmospheric pressure: 70 ~ 106KPaApplicability Micro or rapid Intravenous infusion for various fluid medicine
Electrical Safety Classfication Class I with inner power, IPX4, CF