Infusion Pump WIT-601A

  1. General Details

    More Intelligent : Six Infusion Mode / BOLUS / History Record / Drug Library / Programming Infusion / PIGGYBACK

    More Convenient: Open System / Digital Keyborad and Menu / Flexiable Fixation / Battery discharge time more than 7 hours

    Much Safer: Double CPU / Lock Key / Anti-Reverse Function

    More Comforable: Fluid Warmer Relieve unconfortableness

  2. Features & Specification

    IV Set Options /Size Open System with IV set of 10d/ml、15d/ml、20d/ml、60d/ml(3.4mm~4.5mm); Also can be customized for Dedicated IV set;

    Infusion Mode Volume Mode / Drop Mode / Time Mode /Programming Mode/Weight mode /DrugLiabrary Mode

    Accuracy Error Rate ±5%

    Flow Rate Range 1~1200ml/h(20 d/ml IV set)/ 1~150ml/h(60 d/ml IV set)

    Drip Rate Range 1~400d/min (20d/ml IV set)  / 1~150d/min (60d/ml IV set)

    Preset Volume Range 1-9999ml

    KVO 1ml/h

    Purge 600ml/h

    Bolus Default 600ml/h with volume 2ml;Other value can be setted

    Bubble Detection Detected by ultrasonic with adjustive setting , 50 ~ 800μL.

    Occlusion Pressure  Upper position detected;Down position:High 70-90KPA;Mid 55-70KPA;Low 40-55KPA;

    Alarm Bubbles, Door open, Finished, Pause Overtime,Leakage, Empty , Battery undervoltage or damage, Motor reverse, Incorrect IV Set-type setting, Pressure test failure, Motor failure,Abnormal drip sensor /drip rate, Heater failure, Down Occlusion,Up Occlusion,internal communication fail

    Anti-reversal Infusion Anti-reverse function of motor to prevent upstream

    CPU Dual

    Lock Key Automatic Locking to avoid misoperation.

    Programmable  Infusion can be programmed in 8 segment

    Library User can edit 100 kinds of drug

    History Record 1500 event records , 200 modified recordsRecords can be exported in LAN

    Piggyback User change parameters without switching off

    Fluid Warmer(Selective) Relieve unconfortableness such as pain, vessel convulsion ,shiver when transfuse cold fluid

    Battery Discharge time:Appr. 7 hours ; Recharge time: 12 hours

    Rechargeable Ni-MH battery: DC9.6V, 2000mAh; Vehical-mounted adapter for ambulance(Selective)

    Power Supply AC: 100V ~ 240V, 50Hz/60Hz; Power dissipation :25VA

    Dimension 130mm×120mm×195mm

    Weight 2.0Kg

    Operation Condition Temperature: 5°C ~ 40°C;Relative humidity: 20% ~ 90%
    Atmospheric pressure: 70 ~ 106KPa

    Applicability Micro or rapid Intravenous infusion for various fluid medicine

    Electrical Safety Classfication Class I with inner power, IPX4, CF