Syringe Pump WIT301A
- Features & Specification
Syringe Option and Size Open System with Size:5ml,10ml,20ml,30ml,50ml(& 60ml); Recognized the syringe size automatically;
Injection Mode Rate/Time Mode/Weight mode/Drug Library Mode/Programming Mode
Accuracy Error Rate =< ±2% (mechanical precision=< ±1%)
Flow Rate Range(Increase or decrease: 0.1ml/h)
1.5ml injector: 0.01~150.0ml/h;
2.10ml injector: 0.01~300.0ml/h
3.20ml injector: 0.01~400.0ml/h
4.30ml injector: 0.01~600.0ml/h
5.50ml (& 60ml) injector: 0.01~1200.0ml/hPreset Volume Range 0.1-9999.9ml (0.1ml per step)
KVO 1ml/h
Purge
1.5ml injector: 150.0ml/h
2 .10ml injector: 300.0ml/h;
3.20ml injector: 400.0ml/h;
4 .30ml injector: 600.0ml/h;
5 .50ml (& 60ml) injector: 1200.0ml/hBOLUS Default Rate: 150ml/h, preset value: 2ml, other parameters are selectable.
Occlusion Pressure High: 70~90 Kpa;Mid: 55~70 Kpa;Low: 40~55 KPa
Alarm Near empty , under-voltage battery, power interruption, wrong installation of push-handle clip, wrong installation of compressive bar, wrong installation of pushing bar, end of injection, pause overtime, empty occlusion, abnormal power supply, internal communication fail, abnormal Rate syringe occlusion, battery running-out, battery damage,Rate Overflow, Pressure Sensor Failure, Parameter Error, Pause Over Time, Syringe Size Error
CPU Dual
Lock Key Automatic Locking
Fast Start Overcome dose delay
Back Off ( Anti-Bolus) Overcome the shock dose after occlusion is released
Programmable Injection can be programmed in 8 segment
History Record Inject Record of 1,000 ;Events of 200
Library User can edit up to 100 kinds of drug
Piggyback User can change the Inject Rate without switching off
Battery Battery discharge time:Appr. 7 hours;Recharge time:12 hours; NIMH battery, DC9.6V, 2000mAh; car (adapter) battery: optional.
Power Supply AC100V~240V,50Hz/60Hz ,25VA
Dimension 140mm×145mm×325mm (D x H x L) (Excluding projecting part)
Weight 1.9Kg
Operation Conditions Working temperature: +5℃ ~ +40℃; RH: 20% ~ 90% without condensation; air pressure: 70KPa ~ 106Kpa
Application Applicable for intravenous injection
Electrical Safety Classification Class I, owning internal power supply, IPX4, CF type